Categories

Resuscitator Recall: Mercury Medical Removes Neo-Tee T-Piece Resuscitators due to Risk of Inline Controller Detachment that May Impact Ventilation

Neo-Tee Resuscitators are used for emergency breathing support for neonates and infants. The inline controller may come apart and impact ventilation.

http://www.fda.gov/medical-devices/medical-device-recalls/resuscitator-recall-mercury-medical-removes-neo-tee-t-piece-resuscitators-due-risk-inline-controller

 
Categories

Philips Respironics Issues Additional Usage Instructions for Trilogy Evo Ventilators Related to Use of In-Line Nebulizers

The use of in-line nebulizers placed in certain locations may lead to aerosol deposits forming over time on the ventilator flow sensor. Should this occur, there is a possibility of inaccurate flow measurements affecting therapy.

http://www.fda.gov/safety/recalls-market-withdrawals-safety-alerts/philips-respironics-issues-additional-usage-instructions-trilogy-evo-ventilators-related-use-line